The Pioneer Elite utilizes the 810 nanometer wavelength, the proven technology in dental diode lasers. This wavelength has a high affinity for gingival soft tissue making it an invaluable tool for esthetic sculpting, accessibility, isolation and the healing of soft tissue. The Pioneer Elite comes standard with over 30 pre-set clinical procedures installed along with the ability to easily create your own programs. Included in the software design is the ability to use the Quick Set feature which allows you to simply select Mode, Set Power, and Go! Nothing could be easier.
Request a quote
The Pioneer Elite features technology that provides more convenience, easier operation, more flexibility and unmatched portability. As each clinical situation is unique, the appropriate power setting is primarily determined by the type of lesion, the vascularity of the soft tissue, for example, a more fibrotic tissue requires higher power; or the selected mode of operation, continuous or pulse. The speed of cutting desired by the operator will also determine the power intensity selected, it is always best to use the lowest power setting that will effectively modify the tissue while avoiding tissue charring.
Weight
  • 2.5 lb

Dimensions
  • Unit in cradle: 9.5" x 4" x 6"
  • Unit only: 7" x 4" x 3"
Power range
  • 100mW to 3.0 watts in 100 mW increments
  • Aiming beam (0.3 watts)
Laser classification
  • Laser diode: class 4 laser device

Handpiece sleeves
  • Soft tissue surgery
  • Therapeutics
  • Teeth whitening
Wavelength
  • Laser: 810 nm ± 20 nm
  • Aiming beam: 630 nm - 660 nm ± 15 nm
  • Beam divergence: 9 degrees ± 1 degree
Settings
  • Manual
  • Capability for 8 customized settings
  • 5 Levels of aiming beam intensity
Modes
  • Continuous or pulse mode
  • Hertz rate in pulse mode – fixed: 10 Hz
  • Pulse duration – fixed: .05 Seconds
  • Duty cycle (pulse mode): 50%
Audible and visual notifications
Yes

Power requirements
  • 110-120 VAC @ 60 Hz
  • Wireless pedal – 9V battery
Amperage
3 Amps @ 110-115 VAC

Complies with:
  • IEC 60601-2-22
  • IEC 60825-1
  • 21 CFR 1040.10 and 1040.11
  • UL 60601-1
  • FCC parts 15 & 18 (47 CFR)
  • IC-RSS 210